 The
MP200i series is a 2 wire temperature transmitter with galvanic
isolation between input and output. Micro-processor based, it was
designed for flexibity, accepting mV, Pt100 and a variety of
thermocouples as the input sensor.
Why use isolated
transmitters?
Isolated transmitters protect the electrical
instruments by eliminating ground loop effects and reduce
substantially the undesirable interferences and instabilities in
sensor measurements.
|
 |
 |
- Linearized
Output To Temperature
- High
Quality
- Cold
Junction Compensation for Thermocouples
- Factory
Calibrated
- Isolated
Transmitter
- Proctection
against polarity
inversion
|
 |
- Sensor
input : User defined. The supported sensors are listed in
table 1, along with their maximum ranges.
- Thermocouples
: Types J, K, R, S, T, N, E and B, ccording to IEC 60584 (
ITS-90 ). Impedance >> 1M½
- Pt100
: Excitation : 180 uA. 2 or 3-wire connection (for 2-wire
sensors, tie terminals 2 and 3 together). a = 0.00385,
according to IEC 60751( ITS-90 ).
- Voltage
: 0 a 50mVdc. Impedance >> 1M
- Total
accuracy : better than 0,3% of the maximum range for
thermocouples and 0,2% for
- Pt100
and voltage;
- Response
Time : < 500 ms;
- Output
: 2-wire 4-20mA, linear with respect to the measured
temperature.
- Resolution:
0,001mA ( 14 bits ).
- Power
supply : 12 to 35Vdc, across the transmitter;
- Maximum
load (RL) : RL ( max. )= (V supply – 12) / 0,02 [½]
- Operating
Temperature : -40 to 85°C
- Humidity
: 0 a 90% RH
- Electromagnetic
compatibility : EN 50081-2, EN 50082-2
- Isolation
: 1000Vac for 1 min. between sensor input and the 4-20 mA
loop
| |